Programmable Pulse Generator by Envelope Detection for Implantable Medical Devices

Hao Zhang, Xiaozhou Zhou, Wenglong Zhou

科研成果: 期刊稿件文章同行评审

摘要

Pulse generators in implantable medical devices need to be programmable and miniaturized. However, the existing designs of pulse generator cannot satisfy both of the requirements at the same time. This paper presents a novel design of pulse generators applying magnetic resonance, which is composed of a class-C inverter and an envelope detector, for implantable medical devices. Through simulation and tests, we verify the superiority of our design in programmability of the output pulse signal and miniaturization of the implants, compared with the conventional designs. The amplitude, frequency and duty cycle of the output pulse signal of the implanted receiver can be modulated by controlling the input signal of the transmitter outside the human body. And the footprint of the implanted receiver can be miniaturized to 12 mm × 14 mm × 5 mm, which is smaller than half the size of most of the existing products.

源语言英语
页(从-至)80-89
页数10
期刊IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology
9
1
DOI
出版状态已出版 - 2025