Improved sensitivity of lateral flow assay using paper-based sample concentration technique

Ruihua Tang, Hui Yang, Jane Ru Choi, Yan Gong, Jie Hu, Shangsheng Feng, Belinda Pingguan-Murphy, Qibing Mei, Feng Xu

科研成果: 期刊稿件文章同行评审

89 引用 (Scopus)

摘要

Lateral flow assays (LFAs) hold great promise for point-of-care testing, especially in resource-poor settings. However, the poor sensitivity of LFAs limits their widespread applications. To address this, we developed a novel device by integrating dialysis-based concentration method into LFAs. The device successfully achieved 10-fold signal enhancement in Human Immunodeficiency Virus (HIV) nucleic acid detection with a detection limit of 0.1 nM and 4-fold signal enhancement in myoglobin (MYO) detection with a detection limit of 1.56 ng/mL in less than 25 min. This simple, low-cost and portable integrated device holds great potential for highly sensitive detection of various target analytes for medical diagnostics, food safety analysis and environmental monitoring.

源语言英语
页(从-至)269-276
页数8
期刊Talanta
152
DOI
出版状态已出版 - 15 5月 2016

指纹

探究 'Improved sensitivity of lateral flow assay using paper-based sample concentration technique' 的科研主题。它们共同构成独一无二的指纹。

引用此