[Evaluation of acetabular cup placement precision in Stryker computer-assisted navigated total hip arthroplasty]

Peng Shang, Xueling Bai, Dufang Shi

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

To contrast the methodology of measuring cup placement precision utilizing Mimics and Matlab programming, based on clinical CT images of primary THA cases with computer assisted navigated surgery (CANS) and with the traditional manual method (MANS). The method was applied and analyzed to measure cup anteversion, cup abduction of 50 clinical cases with CANS and MANSThe results show that, cup placement precision differences exits between primary THA cases with CANS and MANS; more cases with CANS are within the safe zone contrasting MANS, and there was less variation and less placement error in CANS cases. CANS can improve cup placement precision and reduce the chance of dislocation efficiently.

Original languageEnglish
Pages (from-to)313-316
Number of pages4
JournalZhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Volume36
Issue number5
StatePublished - 1 Sep 2012
Externally publishedYes

Fingerprint

Dive into the research topics of '[Evaluation of acetabular cup placement precision in Stryker computer-assisted navigated total hip arthroplasty]'. Together they form a unique fingerprint.

Cite this