Establishment and application of new measurement system for work function

Wei Huang, Shan Peng Liu, Hao Zhang, Jin Cao, Wen Qing Zhu, Xue Yin Jiang, Zhi Lin Zhang

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Using the contact potential difference method, we have designed and established a measurement system for work function, which consists of three main parts such as, signal generation part, vibration part and detection part. The reference electrode is vibrated by a low frequency sine wave signal provided by signal generation part. When the output signal of detection part is zero, by changing the load voltage of vibration part, we can get the work function of a sample by calculating the difference between the load voltage and the reference electrode. A change in the work function of ITO is observed when it is treated by UV ozone. The investigations indicate that the work function of ITO has a limiting value and the improvement in the work function of ITO can improve the performance of OLEDs. This system can measure the work function precisely accurately and quickly in atmosphere.

Original languageEnglish
Pages (from-to)454-456+466
JournalGuangdianzi Jiguang/Journal of Optoelectronics Laser
Volume18
Issue number4
StatePublished - Apr 2007
Externally publishedYes

Keywords

  • Contact potential difference
  • ITO
  • OLED
  • Work function

Fingerprint

Dive into the research topics of 'Establishment and application of new measurement system for work function'. Together they form a unique fingerprint.

Cite this