Abstract
Objective To establish a 3D ultrasound imaging system based on pulse-triggered image acquisition using the linear probe on the VerasonicsTM vantage 128 platform and evaluate its performance in scanning standard phantom and human carotid artery. Methods The 3D ultrasound imaging system included 3 modules for probe motion control, image acquisition and storage, and 3D image reconstruction and display. To improve the precision of image acquisition, we used fixed frequency pulses to control the external trigger function combined with mechanical scanning. Voxel-based 3D reconstruction was used for image reconstruction and display. The user interface was designed to allow direct operations of the platform. We carried out scanning tests of standard ultrasound phantom and human carotid artery to evaluate the performance of this imaging system. Results We successfully constructed a 3D ultrasound imaging system based on pulse-triggered image acquisition. The results of standard phantom and human carotid scanning tests showed that each module of the system was fully functional. The self-designed user interface of this ultrasound imaging system allowed full control of the system functions for original image acquisition, 3D image reconstruction, and display of cross-sections in 3 different views. Conclusion This 3D ultrasound imaging system achieves high-quality 3D ultrasound imaging and provides the basis for further study and clinical application of 3D ultrasound imaging.
Translated title of the contribution | Establishment of a 3D ultrasound imaging system based on pulse-triggered image acquisition |
---|---|
Original language | Chinese (Traditional) |
Pages (from-to) | 767-774 |
Number of pages | 8 |
Journal | Nan Fang Yi Ke Da Xue Xue Bao / Journal of Southern Medical University |
Volume | 41 |
Issue number | 5 |
DOIs | |
State | Published - 2021 |