T1 - The effect of subcutaneous fat on electrical impedance myography
T2 - Electrode configuration and multi-frequency analyses
AU - Li, Le
AU - Li, Xiaoyan
AU - Hu, Huijing
AU - Shin, Henry
AU - Zhou, Ping
N1 - Publisher Copyright:
© 2016 Li et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2016/5/1
Y1 - 2016/5/1
N2 - This study investigates the impact of the subcutaneous fat layer (SFL) thickness on localized electrical impedance myography (EIM), as well as the effects of different current electrodes, varying in distance and direction, on EIM output. Twenty-three healthy subjects underwent localized multi-frequency EIM on their biceps brachii muscles with a hand-held electrode array. The EIM measurements were recorded under three different configurations: wide (or outer) longitudinal configuration 6.8 cm, narrow (or inner) longitudinal configuration 4.5 cm, and narrow transverse configuration 4.5 cm. Ultrasound was applied to measure the SFL thickness. Coefficients of determination (R2) of three EIM variables (resistance, reactance, and phase) and SFL thickness were calculated. For the longitudinal configuration, the wide distance could reduce the effects of the subcutaneous fat when compared with the narrow distance, but a significant correlation still remained for all three EIM parameters. However, there was no significant correlation between SFL thickness and reactance in the transverse configuration (R2 = 0.0294, p = 0.434). Utilizing a ratio of 50kHz/100kHz phase was found to be able to help reduce the correlation with SFL thickness for all the three configurations. The findings indicate that the appropriate selection of the current electrode distance, direction and the multi-frequency phase ratio can reduce the impact of subcutaneous fat on EIM. These settings should be evaluated for future clinical studies using hand-held localized arrays to perform EIM.
AB - This study investigates the impact of the subcutaneous fat layer (SFL) thickness on localized electrical impedance myography (EIM), as well as the effects of different current electrodes, varying in distance and direction, on EIM output. Twenty-three healthy subjects underwent localized multi-frequency EIM on their biceps brachii muscles with a hand-held electrode array. The EIM measurements were recorded under three different configurations: wide (or outer) longitudinal configuration 6.8 cm, narrow (or inner) longitudinal configuration 4.5 cm, and narrow transverse configuration 4.5 cm. Ultrasound was applied to measure the SFL thickness. Coefficients of determination (R2) of three EIM variables (resistance, reactance, and phase) and SFL thickness were calculated. For the longitudinal configuration, the wide distance could reduce the effects of the subcutaneous fat when compared with the narrow distance, but a significant correlation still remained for all three EIM parameters. However, there was no significant correlation between SFL thickness and reactance in the transverse configuration (R2 = 0.0294, p = 0.434). Utilizing a ratio of 50kHz/100kHz phase was found to be able to help reduce the correlation with SFL thickness for all the three configurations. The findings indicate that the appropriate selection of the current electrode distance, direction and the multi-frequency phase ratio can reduce the impact of subcutaneous fat on EIM. These settings should be evaluated for future clinical studies using hand-held localized arrays to perform EIM.
UR - http://www.scopus.com/inward/record.url?scp=84973305008&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0156154
DO - 10.1371/journal.pone.0156154
M3 - 文章
C2 - 27227876
AN - SCOPUS:84973305008
SN - 1932-6203
VL - 11
JO - PLoS ONE
JF - PLoS ONE
IS - 5
M1 - e0156154
ER -