Rationale of Time-to-Count technology to measure radiant intensity

Zhi Qiang Liu, Hong Guang Ma, Wen Yan Ma, Guo Hua Wang, Deng He Liu

科研成果: 期刊稿件文章同行评审

2 引用 (Scopus)

摘要

GM tube is widely used in the field of radiation measurement. But it's testing accuracy is low. Because the traditional measuring method is affected by the die-time of GM tube. This paper introduces the Time-to-Count technology, which is not affected by die-time and may enhance the testing accuracy of GM tube, and theoretically proves the mathematic rationale of this method. Thereafter, a new method for estimating radiant intensity is proposed, and the new estimate formula is proved to be an unbiased and asymptotic-optimum estimator, and the best estimator in the form of a/∑ti. At last the correctness of this method is verified by computer simulation.

源语言英语
页(从-至)41-44+125
期刊Hedianzixue Yu Tance Jishu/Nuclear Electronics and Detection Technology
26
1
出版状态已出版 - 1月 2006
已对外发布

指纹

探究 'Rationale of Time-to-Count technology to measure radiant intensity' 的科研主题。它们共同构成独一无二的指纹。

引用此