T1 - Chip-based ion chromatography (chip-IC) with a sensitive five-electrode conductivity detector for the simultaneous detection of multiple ions in drinking water
AU - Li, Xiaoping
AU - Chang, Honglong
N1 - Publisher Copyright:
© 2020, The Author(s).
PY - 2020/12/1
Y1 - 2020/12/1
N2 - The emerging need for accurate, efficient, inexpensive, and multiparameter monitoring of water quality has led to interest in the miniaturization of benchtop chromatography systems. This paper reports a chip-based ion chromatography (chip-IC) system in which the microvalves, sample channel, packed column, and conductivity detector are all integrated on a polymethylmethacrylate (PMMA) chip. A laser-based bonding technique was developed to guarantee simultaneous robust sealing between the homogeneous and heterogeneous interfaces. A five-electrode-based conductivity detector was presented to improve the sensitivity for nonsuppressed anion detection. Common anions (F−, Cl−, NO3−, and SO42−) were separated in less than 8 min, and a detection limit (LOD) of 0.6 mg L−1 was achieved for SO42−. Tap water was also analyzed using the proposed chip-IC system, and the relative deviations of the quantified concentration were less than 10% when compared with that a commercial IC system.
AB - The emerging need for accurate, efficient, inexpensive, and multiparameter monitoring of water quality has led to interest in the miniaturization of benchtop chromatography systems. This paper reports a chip-based ion chromatography (chip-IC) system in which the microvalves, sample channel, packed column, and conductivity detector are all integrated on a polymethylmethacrylate (PMMA) chip. A laser-based bonding technique was developed to guarantee simultaneous robust sealing between the homogeneous and heterogeneous interfaces. A five-electrode-based conductivity detector was presented to improve the sensitivity for nonsuppressed anion detection. Common anions (F−, Cl−, NO3−, and SO42−) were separated in less than 8 min, and a detection limit (LOD) of 0.6 mg L−1 was achieved for SO42−. Tap water was also analyzed using the proposed chip-IC system, and the relative deviations of the quantified concentration were less than 10% when compared with that a commercial IC system.
UR - http://www.scopus.com/inward/record.url?scp=85089729737&partnerID=8YFLogxK
U2 - 10.1038/s41378-020-0175-x
DO - 10.1038/s41378-020-0175-x
M3 - 文章
AN - SCOPUS:85089729737
SN - 2055-7434
VL - 6
JO - Microsystems and Nanoengineering
JF - Microsystems and Nanoengineering
IS - 1
M1 - 66
ER -