摘要
A Selenium thin film transistor (STFT) array-based rotational volume tomographic digital angiography (VTDA) imaging system has been constructed. The system consists of an x-ray tube and a STFT detector that are separately mounted on a gantry. This system uses the STFT array as a two-dimensional (2D) detector so that a set of 2D projections can be acquired for a direct three-dimensional (3D) reconstruction. This paper presents the results of the preliminary phantom studies using the STFT-based volume tomographic angiography imaging system. This research work demonstrates through phantom studies that the STFT array introduces no distortion and the STFT-based VTDA imaging system has a higher spatial resolution and better contrast resolution than an image intensifier(II)-based system.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 316-324 |
| 页数 | 9 |
| 期刊 | Proceedings of SPIE - The International Society for Optical Engineering |
| 卷 | 3336 |
| DOI | |
| 出版状态 | 已出版 - 1998 |
| 已对外发布 | 是 |
| 活动 | Medical Imaging 1998: Physics of Medical Imaging - San Diego, CA, 美国 期限: 22 2月 1998 → 24 2月 1998 |