Relative and Absolute Interrater Reliabilities of a Hand-Held Myotonometer to Quantify Mechanical Muscle Properties in Patients with Acute Stroke in an Inpatient Ward

  • Wai Leung Ambrose Lo
  • , Jiang Li Zhao
  • , Le Li
  • , Yu Rong Mao
  • , Dong Feng Huang
  • Sun Yat-Sen University
  • University College London

科研成果: 期刊稿件文章同行评审

53 引用 (Scopus)

摘要

Introduction. The reliability of using MyotonPRO to quantify muscles mechanical properties in a ward setting for the acute stroke population remains unknown. Aims. To investigate the within-session relative and absolute interrater reliability of MyotonPRO. Methods. Mechanical properties of biceps brachii, brachioradialis, rectus femoris, and tibialis anterior were recorded at bedside. Participants were within 1 month of the first occurrence of stroke. Relative reliability was assessed by intraclass correlation coefficient (ICC). Absolute reliability was assessed by standard error of measurement (SEM), SEM%, smallest real difference (SRD), SRD%, and the Bland-Altman 95% limits of agreement. Results. ICCs of all studied muscles ranged between 0.63 and 0.97. The SEM of all muscles ranged within 0.30-0.88 Hz for tone, 0.07-0.19 for decrement, 6.42-20.20 N/m for stiffness, and 0.04-0.07 for creep. The SRD of all muscles ranged within 0.70-2.05 Hz for tone, 0.16-0.45 for decrement, 14.98-47.15 N/m for stiffness, and 0.09-0.17 for creep. Conclusions. MyotonPRO demonstrated acceptable relative and absolute reliability in a ward setting for patients with acute stroke. However, results must be interpreted with caution, due to the varying level of consistency between different muscles, as well as between different parameters within a muscle.

源语言英语
期刊论文编号4294028
期刊BioMed Research International
2017
DOI
出版状态已出版 - 2017
已对外发布

学术指纹

探究 'Relative and Absolute Interrater Reliabilities of a Hand-Held Myotonometer to Quantify Mechanical Muscle Properties in Patients with Acute Stroke in an Inpatient Ward' 的科研主题。它们共同构成独一无二的学术指纹。

引用此