A transverse pressure sensor with high-sensitivity based on a mechanically induced long-period fiber grating (MLPFG) and fiber loop ring-down technique is presented. When a MLPFG is spliced into a fiber loop, an extra loss is introduced, which leads to a decrease of the ring-down time. The results demonstrate that the difference between the reciprocals of the ring-down time with and without pressure increases exponentially with increasing the pressure in the range of 0-23.4 MPa. This sensor shows good repeatability, and the least detectable pressure is only 0.0068 MPa which is about 18 times less than detecting the output light intensity directly.