摘要
Quasi-hemispherical CdZnTe detector was manufactured successfully to fully understand the performance in the mixed gamma–neutron detection field. Together with the software of COMSOL, Geant4, and Matlab, the detector structure has been optimized. The CdZnTe detector performs good energy resolutions for 241 Am, 57 Co, and 137 Cs radiation sources, especially for 137 Cs (10.91 keV full width at half maximum [FWHM] at 662 keV). A linear relationship between the energy positions and spectrum channels indicates that the detector is effective for the precise energy detection from 59.5 to 662 keV. Finally, neutron and gamma events were detected simultaneously at room temperature using 241 AmBe neutron source. The spectrum shows good energy resolution for neutron capture gamma ray (14.28 keV FWHM at 558 keV). Our work demonstrates that the quasi-hemispherical CdZnTe detector is promising for simultaneous detection of neutrons and gamma radiation.