摘要
A technique is described to remove the coaxial-to-waveguide adaptor errors when the reflection coefficient of waveguide device is measured by vector network analyzer. This technique utilizes a matched load and an adjustable shorter with several unspecified phases. And one of the shorted planes is used as reference plane. It is shown that the technique is resistant to the error of the adjustable shorter position and is easy to implement. The efficiency of this technique is verified by measurement data in X band.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 168-171 |
| 页数 | 4 |
| 期刊 | Jiliang Xuebao/Acta Metrologica Sinica |
| 卷 | 32 |
| 期 | 2 |
| DOI | |
| 出版状态 | 已出版 - 3月 2011 |
指纹
探究 'A coaxial-to-waveguide adaptor calibration method using an adjustable shorter with unspecified phases' 的科研主题。它们共同构成独一无二的指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver