A porous bioactive titania coating on biomedical β titanium alloy was prepared by pre-anodization followed by micro arc oxidation technology. The effects of pre-anodization on the phase constituent, morphology and electrochemical corrosion behavior of the microarc oxidation coating were investigated. The results show that pre-anodization has less influence on the phase constituent and the surface morphology of the microarc oxidation coating, but improves the inner layer density of the microarc oxidation coating. The decrease of plasma discharge strength due to the presence of the pre-anodized oxide film contributes to the formation of the compact inner layer. The pre-anodized microarc oxidation coating effectively inhibits the penetration of the electrolyte in 0.9% NaCl solution and thus increases the corrosion resistance of the coated titanium alloy in physiological solution.

Original languageEnglish
Pages (from-to)3775-3779
Number of pages5
JournalMaterials Science and Engineering C
Volume33
Issue number7
DOIs
StatePublished - Oct 2013
Externally publishedYes

Keywords

  • Coating
  • Corrosion
  • Electrochemistry
  • Titanium alloy