Abstract
γ-ray spectroscopy response for CdZnTe planar detector (10 mm×10 mm×2.5 mm) made by ourself is evaluated using an un-collimated 241Am (59.5 keV) radioactive source at room temperature for 10 h continuously, in which the polarization is not occurred. The γ-ray (59.5 keV) response spectra of the detector are acquired for a long time (1 year), it shows that operation of the detector is stable. A thick CdZnTe capacitive Frisch grid detector (10 mm×5 mm×10 mm) is fabricated in order to approach high collection efficiency for high-energy γ-rays. An excellent energy resolution of 3.65% for 137Cs at 662 keV is obtained without any additional signal processing.
| Original language | English |
|---|---|
| Pages (from-to) | 331-334 |
| Number of pages | 4 |
| Journal | Guangdianzi Jiguang/Journal of Optoelectronics Laser |
| Volume | 21 |
| Issue number | 3 |
| State | Published - Mar 2010 |
Keywords
- γ-ray
- CdZnTe detector
- Electron mobility life-time product
- Polarization
Fingerprint
Dive into the research topics of 'Investigation on γ-ray spectroscopy response and stability for CdZnTe detectors'. Together they form a unique fingerprint.Cite this