A novel electrochemical biosensor for detection of cholesterol

Lin Xu, Yiting Hou, Mengdan Zhang, Xin Yang, Greath Jenkins, Wei Huang, Cheng Yao, Qiong Wu

科研成果: 期刊稿件文章同行评审

12 引用 (Scopus)

摘要

We report on a highly sensitive electrochemical biosensor for determination of cholesterol. The biosensor was fabricated by co-immobilizing bi-enzymes, cholesterol oxidase (ChOx), and horseradish peroxidase (HRP). Voltammetric technique such as cyclic voltammetry and impedance experiment were used to study the characterization of modified electrode step by step. The developed sensor is cheap, disposable, portable and exhibits higher sensitivity. The biosensor expressed a wide linear range up to 300 mg dL–1 in a physiological condition (pH 7.0), with a correlation coefficient of 0.9969. A sensitivity of 13.28 μA mg–1 dL cm−2 which makes it very promising for the clinical determination of cholesterol.

源语言英语
页(从-至)239-244
页数6
期刊Russian Journal of Electrochemistry
52
3
DOI
出版状态已出版 - 1 3月 2016
已对外发布

指纹

探究 'A novel electrochemical biosensor for detection of cholesterol' 的科研主题。它们共同构成独一无二的指纹。

引用此